Cubic measuring instrument

 Nothing

详情介绍

1 host model Royal 575

2. Probe seat MH20I manual double rotation indexing probe seat of probe system

Probe module tp20-std

See needle list for needle test

Calibration tee

Calibration ball phi is 25

3. Control system RENISHAW UCC

4 computer system

Computer dual core 3.3g /4G/500G

Printer A4 inkjet printer

5 software system Rational DMIS

Technical data installation and maintenance manual

Operation manual

Certificate of product delivery certificate

8 quick plug 10

Tracheal phi 10

 

B. overview of host:

 

Royal series CMM is the latest generation CMM of nano precision. It adopts the international advanced finite element analysis method to design the precision beam technology. The design adopts the rigid and light moving bridge structure of the closed frame. Equipped with the internationally renowned brand high-performance special control system for three coordinates and the internationally advanced Rational DMIS measurement software package, it has become a relatively high cost performance product among similar models.

 

 

C. Introduction to the host:

 

01. Model name: Roya 575 CMM

 

02. Measuring range (mm) : X*Y*Z=500×700×500

 

03. Appearance size (mm) : X*Y*Z=1678×2260×2900

 

Weight (kg) : 2925kg

 

Maximum load (kg) : 1300kg

 

 

B. technical description

 

A. host:

 

1. Royal series measuring machine adopts the internationally advanced finite element analysis method to design the precision rectangular beam technology, and the rectangular beam adopts the air bearing surrounded by four sides and the dust-proof design and technical characteristics of organ type.

 

2. The whole mechanical structure adopts the fully closed frame moving bridge structure with good rigid structure and light weight. Its structure is simple, compact, the bearing capacity is large, the movement performance is good.

 

3. Fix the high-quality granite worktable, so that it has strong bearing capacity, wide and convenient loading and unloading space.

 

4, y-direction guide adopts dovetail type, high positioning accuracy, good stability.

 

5, the triaxial use of high-quality granite, thermal expansion coefficient is small, the three axes have the same temperature characteristics, so it has good temperature stability, anti-actual deformation ability, good rigidity, dynamic geometric error deformation is small.

 

6, the three axes are self-cleaning pre-loaded high-precision air bearings composed of static pressure air floating guide rail, the bearing span is large, anti-angle pendulum ability is strong, resistance is small, no wear, more stable movement.

 

7, the imported reflective metal strip grating metering system, and most of the workpiece has a similar coefficient of thermal expansion, to ensure a good repeatability.

 

8. The triaxial drive system adopts unique patented technology (patent no. : ZL 2014 2 0529248.2) and wire rope transmission, which makes the following error of the system smaller and the response speed of the system faster in the process of movement, greatly improves the dynamic performance of large measuring equipment, and minimizes the impact of motion inertia on the measurement accuracy. The application of wire rope transmission makes the use and maintenance of the measuring machine cheaper and more convenient.

 

9. Z axis adopts cylinder balancing device, which greatly improves the positioning accuracy and stability of Z axis.

 

10. The control system adopts the British RENISHAW ucc system, which has the internationally advanced double-computer system of upper and lower position machines, thus greatly improving the reliability and anti-interference ability of the system and reducing the maintenance cost.

11. The software adopts the powerful American Rational DMIS measurement software package, and its perfect measurement function and online function provide the perfect measurement solution for users.

 

B. probe system:

 

01, probe

 

RENISHAW MH20I manual double rotation indexing head seat

 

Special point: manual double rotation measuring head seat, can realize the space 15° step any Angle rotation.

 

Technical parameters: weight: 210g (7.4 oz)

 

Interface: TP20

 

Rotation and locking: manual biaxial locking, clockwise locking, inverse

 

Turn the hour hand loose.

 

Indexing: A Angle 0° ~ 90°, 15° step distance

 

Angle B 0° ~ ±180°, 15° step distance

 

Test seat status: 1 LED indicator light

 

Maximum load: EM2 (94.5mm) module

 

 

 

RENISHAW TP20 standard force measurement module

 

Technical parameters: induction direction ±X ±Y + Z

 

Needle measuring interface M2

 

22 g in weight

 

Rated life > 106 times

 

 

 

02, calibrator: standard ball Φ 25 and 1 set

 

03. Precision index:

 

Length measuring system: imported from the UK RENISHAW precision grating ruler

 

RENISHAW head resolution: 0.5 um

 

Indication error: E≤1.2+ 3.0l /1000 um

 

Detection error: R≤ 1.2um

 

C. environmental requirements

 

Temperature: 20+2℃ 1℃/h 1℃/m 2℃/24h

 

Humidity: 40% to 70%

 

Power supply: 220V±10%, 50hz-60hz

 

Source: acuity 0.55 Mpa